Digiducer 333D01 USB set (Code: 47001)

Digiducer USB set

This package includes digiducer 333D01 sensor and the analysis software Akulap vibro. A computer is not included.

Digiducer's Model 333D01 USB Digital Accelerometer simplifies vibration measurements.

The Model 333D01 is the first fully integrated and ruggedized packaging of a high-resolution, broad-frequency piezoelectric accelerometer with integrated internal digital data acquisition. Taking reliable, quality vibration measurements is as simple as recording audio input in a tablet or PC.  The 333D01 Piezolelectric USB Digital Accelerometer serves as a new high fidelity tool for the test, industrial and app specific markets.

Specification USB sensor

  • Piezoelectric ruggedness and extended dynamic range
  • Eliminates need for data acquisition
  • Flat response up to 8 kHz
  • 24 bit internal ADC
  • Record vibration data via laptop or tablet

Specification Software

  • High resolution FFT Analyzer <0.003Hz
  • permanent calibration
  • Bandwith DC to 20kHz
  • Rich set of DSP functions, weighting, filtering, trigger etc.
  • requires windows XP,vista,7,8,10 32/64bit

Specification USB sensor

Sensitivity
    

Channel A:             4.3 % FSV/g
Channel B:             8.4 % FSV/g
    

Channel A:             0.44 % FSV/m/s2
Channel B:             0.86 % FSV/m/s2
Measurement Range3     

Channel A:                  ± 20 g pk
Channel B:                  ± 10 g pk
    

Channel A:                       ± 196 m/s2
Channel B:                         ± 98 m/s2
ADC Bandwidth (-3dB)     9.3 to 137400 cpm     0.155 to 22900 Hz
Frequency Range (±5%)3     120 to 480000 cpm     2 to 8000 Hz
Frequency Range (±10%)3     90 to 660000 cpm     1.5 to 11000 Hz
Frequency Range (±3dB)3     54 to 900000 cpm     0.9 to 15000 Hz
Resonant Frequency     ≥ 1500000 cpm     ≥ 25000 Hz
Mounted Resonance3     1044000 cpm     17400 Hz
Mounted Resonance Amplification3     200 %     200 %
Broadband Resolution1 (1 Hz to 10,000 Hz)     0.0025 g pk     0.0245 m/s2 pk
Non-Linearity4     ≤ 2 %     ≤ 2 %
Transverse Sensitivity3     ≤ 5 %     ≤ 5 %

 
Environmental
Overload Limit (Shock)     7000 g pk     68647 m/s2 pk
Temperature Range     14 to 158 °F     -10 to +70 °C
Temperature Coefficient     0.10 % / °F     0.18 % / °C

 
Electrical
Communication Standard     USB 2.0 Full Speed     USB 2.0 Full Speed
Power Consumption3     ≤ 45 mA     ≤ 45mA
Internal ADC     24 bit     24 bit
Supported Sample Rates     

24 bit @ 48, 44.1, 32, 22.05,
16, 11.025, 8 kHz
16 bit @ 48, 44.1, 32, 22.05,
16, 11.025, 8 kHz
    

24 bit @ 48, 44.1, 32, 22.05,
16, 11.025, 8 kHz
16 bit @ 48, 44.1, 32, 22.05,
16, 11.025, 8 kHz

 
Mechanical
Size ─ Hex     1.0 in     25.4 mm
Size ─ Height     2.6 in     66.0 mm
Weight     4.62 oz     131 gram
Mounting Thread     1/4-28 UNF     1/4-28 UNF
Mounting Torque     2 to 5 lbf·ft     2.7 to 6.8 N·m
Sensing Element     Piezoelectric Ceramic     Piezoelectric Ceramic
Sensing Geometry     Shear     Shear
Housing Material     Stainless Steel     Stainless Steel
Sealing     Welded Hermetic     Welded Hermetic
Electrical Connector     USB Type A Male     USB Type A Male
Electrical Connection Position     Top     Top
Cable (Integral) Length     9.6 ft     2.9 m