AUDIO PRECISION APx555 • REFERENZLABOR DRESDEN

Audio Precision APx555: Reference Analyzer for Contract Testing, Calibration & Psychoacoustics

The Audio Precision APx555 is the globally recognized reference standard in electronic and electroacoustic metrology. In our Dresden laboratory, it serves as the master instrument for traceable microphone and calibrator verifications, independent B2B contract testing of your hardware, and our native psychoacoustic plugins for APx Sequence Mode.

Technical Specifications of Our APx555
Residual THD+N
< -120 dB (0,0001 %)
Analog Bandwidth
DC – 1 MHz
Residual Input Noise
≤ 1,0 µVrms
FFT Resolution
2,5 Mio. Punkte
Verified via documented loopback self-test (balanced XLR, 1 kHz at 2.0 Vrms, amplitude flatness ±0.008 dB).
1. Physical Architecture • High-Performance Sine Analyzer (HPSA)

How Do You Measure -120 dB THD+N When the Best 24-Bit ADCs Physically Max Out at -115 dB?

Modern A/D and D/A converters outperform vintage analog gear in almost every metric. Yet any purely digital converter chain hits a physical ceiling around -110 dB to -115 dB due to quantization noise, clock jitter, and reconstruction filters. The APx555 therefore combines a microprocessor-controlled analog RC oscillator with a dual-path twin-T notch receiver.

1. Hybrid Generator RC-Oszillator vs. DAC 0,001 Hz – 204 kHz THD+N < -120 dB 2,00 Vrms @ 1 kHz 2. Input Ranging Relais-Teiler (ab 1,0 V) Optimum: ≥ 1,0 Vrms BNC / XLR Symmetrisch Eigenrauschen ≤ 1,0 µV Path A: Level ADC Pegelreferenz Grundwelle DC – 1 MHz Bandbreite Path B: Analog Twin-T Notch 1 kHz Grundton unterdrückt +60 dB Verstärkung Residuum Schützt ADC vor Übersteuerung 4. DSP & FFT Dual-Path-Rekonstruktion -120,4 dB 0,000095 % THD+N 2,5 Mio. Punkte FFT
Interactive HPSA Architecture & Ranging Simulator
Analyzer Input Voltage (1 kHz Loopback): 2,00 Vrms
0,1 V (Rauschdominiert) 1,0 V (ADC-Dämpfungsglied-Schwelle) 5,0 V
Analog 1 kHz Twin-T Notch Filter (Fundamental Rejection before ADC):
Resulting Loopback Measurement
Measured THD+N (dB)
-120,4 dB
Distortion THD+N (%)
0,000095 %

At 1.0 Vrms and above, the APx555 programmable input attenuator operates in its optimal window. The analog notch filter strips the 1 kHz fundamental prior to the ADC, amplifies the residual harmonics by 1000x (+60 dB), and achieves below -120 dB (0.000095 %) THD+N.

Engineering Note (Gain Relays & Shielding): When measuring below 0.00005 %, proper grounding, magnetic shielding, and the switching thresholds of the internal input relays (e.g., at 1.0 V) are decisive. Some audio manufacturers intentionally tune output stages just below these relay steps to game datasheet numbers—we sweep across the entire level range (THD+N vs. Level).
2. Lab Testing Service • Independent B2B Contract Measurements

Why Qualify Your Electronics with Basic Audio Interfaces? Use Our APx555 Lab.

Not every R&D department needs to purchase a €50,000 analyzer for occasional verification. We test your prototypes, preamplifiers, ADCs/DACs, transformers, and power amplifiers (up to 2 kW) confidentially at our Dresden lab—either as a turnkey test report or via live remote access.

Turnkey Lab Report #9012

Measurements in our lab with Audio Precision APX555

You ship your Device Under Test (DUT) to Dresden. We configure the shielded test setup, execute the measurements on our Audio Precision APx555, and return your hardware with a comprehensive PDF test report.

  • • Covers 5 standardized parameters across 2 channels (e.g., frequency response, THD+N vs. level, THD+N vs. frequency, SNR, crosstalk/FFT)
  • • Analog and digital interfaces (XLR/BNC balanced/unbalanced, AES/EBU, S/PDIF, ASIO)
  • • 100% confidential execution for B2B R&D projects (no publication of your data)
199,00 € Net
236,81 € incl. 19% VAT • net plus shipping
Remote via TeamViewer #9012DIY

DIY: Measurements in our lab with Audio Precision APX555 via remote

For hardware engineers already familiar with the APx software suite who want to run custom test sequences interactively on their device.

  • • We wire your shipped DUT to the APx555 strictly according to your specifications (hardware setup included)
  • • 1 hour of exclusive real-time remote access via TeamViewer to our laboratory analyzer
  • • Direct export of your APx project files, raw CSV data, and PDF plots
80,00 € Net
95,20 € incl. 19% VAT • net plus shipping
Custom & High-Power #9011

Measurements in our lab with e.g. Audio Precision APX555

Custom test benches for power amplifiers up to 2 kW into dummy loads, ultrasonic sensors up to 1 MHz, Keysight/Picoscope setups, or on-site testing.

  • • Power amplifier testing up to 2 kW (Class-D carrier analysis up to 1 MHz, 19/20 kHz IMD, 32-tone multitone)
  • • Supplementary lab fleet: Keysight U8903A, Keysight 33500B, Rigol DM3061 (6.5 digits), G.R.A.S. 40HL (5.5 dB(A))
  • • Combinable with acoustic testing in our Spektra SQ-4.2 precision pressure chamber (up to 170 dB SPL)
Price on request
Custom quotation based on test specification (plus shipping)
3. In-House Software Engineering • Psychoacoustic Plugins for Audio Precision APx

Psychoacoustics Directly in APx Sequence Mode: Loudness in Sone (DIN 45631/A1), Sharpness & Tonality

Audio Precision's APx software supports the APx555, APx515, APx525, APx585, and APx Flex series. Via the Sequence Mode plug-in interface, our software modules embed standards-compliant psychoacoustic metrics directly into your automated R&D and end-of-line test sequences.

Seamless Integration of Akulap into the Audio Precision APx Framework

Simple A-weighted sound pressure levels in dB(A) fail to capture human annoyance perception for fans, HVAC units, automotive actuators, or IT equipment. Our APx plugins pass the time-domain signals acquired by the APx analyzer directly to our Akulap calculation engine in the background and return the psychoacoustic metrics straight into your APx test report.

Psychoacoustic Parameter Applicable Standard / Method APx Plug-in SKU
Time-Varying Loudness N(t) & N5 in Sone DIN 45631/A1 • ISO 532-1 (Zwicker) #75003 & #75004
Stationary Loudness & Specific Loudness DIN 45631 • ISO 532B #75004 (Mehrkanal)
Sharpness in acum DIN 45692 #75004
Tonality / Tone Penalty KT DIN 45681:2006 #75004
Roughness (Daniel & Weber) & Impulsiveness Daniel & Weber Modell #75004
Don't Own an Audio Precision Analyzer?
All psychoacoustic calculation models are also available as a standalone PC measurement system in Akulap.
View Loudness & Sone Portal →
APx Sequence Mode Plug-in #75003

time variant loudness sone DIN45631/A1

Calculates time-varying loudness in Sone per DIN 45631/A1 and ISO 532-1 directly inside Sequence Mode for APx555, APx515, APx525, APx585, and APx Flex. Includes USB dongle, manual, and test vectors.

APx Complete Suite #75004

Complete psycho-acoustic measurement suite for the APX framework

Complete multi-channel psychoacoustic suite for the APx framework: Stationary and time-varying loudness (ISO 532B / ISO 532-1), specific loudness, sharpness (DIN 45692), tonality (DIN 45681), roughness, impulsiveness, percentiles, and spectrogram.

Standalone Windows PC Alternative (No APx Hardware Required) #3303026
Plug-in Psychoacoustics II
2420,00 € Net 2879,80 € incl. 19% VAT • net plus shipping
4. Public Bench-Test Archive • Real-World APx555 Measurements

From Our Dresden Test Lab: 9 Commercial Audio Devices Bench-Tested on the APx555

Manufacturer datasheets frequently omit behavior at clipping boundaries or low frequencies. Below is a selection of our published laboratory investigations:

Mic Booster (Report #196) +26 dB Gain • +6 dB SNR

Klark Teknik MicBooster CM-1 on the APx555: Cleaner Than the Analyzer Input?

At microvolt levels (50 µV), the P48V-powered CM-1 lowers the effective noise floor of the measurement chain by 6 dB while maintaining a flat response up to 80 kHz.

Verdict: Exceptional noise performance for low-output dynamic mics.
Read Full Lab Report with APx555 Plots →
Comparison (Report #197) -45 dB THD+N • 7 kHz Schwingen

Klark Teknik CM-1 vs. Two €5 Low-Cost Import Clones

Instead of the claimed 25 dB, the ultra-budget clone produced 38 dB gain alongside a parasitic 7 kHz oscillation, severe 50 Hz mains hum, and -45 dB THD+N.

Verdict: Unusable for measurement or studio applications.
Read Full Lab Report with APx555 Plots →
Preamplifier (Report #113) 86 dB SNR (Z) • Netz vs. Akku

IMG Stageline MPA-102 Mic Preamp: AC Mains vs. 9V Battery Operation

Full characterization of the 12 switched gain steps (15–70 dB), the 100 Hz / 12 kHz filters, and 50 Hz mains hum ingress on AC power versus battery power.

Verdict: Solid field preamp achieving 86 dB unweighted SNR.
Read Full Lab Report with APx555 Plots →
Class-D Amplifier (Report #172) 2x 100 W @ 4 Ω • TPA3255

Aiyima A07 Class-D Stereo Power Amplifier (TI TPA3255) on the APx555

Evaluation of frequency response, THD+N versus output power up to 2x 100 W into 4 ohms, 32-tone multitone spectrum, and 19/20 kHz intermodulation distortion.

Verdict: High power density suitable for compact acoustic test rigs.
Read Full Lab Report with APx555 Plots →
RC Oscillator (Report #118) 0,0063 % THD vs. 0,0001 %

Testing a Commercial 'Ultra-Low Distortion' RC Sine Generator

Purely analog RC oscillators can beat digital DACs—but require extreme circuit precision. This commercial unit measured 0.006311 % THD (0.00754 % THD+N) at 1 kHz.

Verdict: 63x higher distortion than the internal APx555 RC generator.
Read Full Lab Report with APx555 Plots →
Transformer 1:4 (Report #114) Linear: 1 mV – 20 mV

Neutrik NTE-4 Audio Transformer (1:4): Core Saturation & 40 Hz THD

Between 1 mV and 20 mV thermal noise dominates; above 20 mV magnetic core non-linearities rise sharply, particularly at low frequencies (40 Hz).

Verdict: Low-distortion operation is restricted to small signals below 20 mV.
Read Full Lab Report with APx555 Plots →
Transformers 1:1 (Reports #115 & #117) 600 Ω • 40-Hz-Spektralanalyse

Neutrik NTL-1 (1:1) & Welter 1:1 Small-Signal Transformers Compared

Detailed comparison of frequency response, odd-order harmonics (H3, H5) at 40 Hz, and the usable dynamic window between noise floor and core saturation.

Verdict: Precise metrological proof of magnetic saturation thresholds.
Read Full Lab Report with APx555 Plots →
Wireless Link (Report #164) 6,8 ms Delay • 260 mV Clip

Lekato 5.8 GHz XLR Wireless System: Latency, Clipping & Two-Tone Artifacts

APx555 testing revealed 95 dB(A) SNR and 6.8 ms latency, but hard clipping above 260 mVrms and severe data-compression artifacts under two-tone excitation.

Verdict: Unsuitable for acoustic metrology due to digital compression.
Read Full Lab Report with APx555 Plots →
Mic Teardown (Report #143) 500 mV vs. 10 Vrms (28 Vpp)

Electronics Teardown: Behringer ECM8000 & MiniDSP UMIK-1/2 vs. ATD5-T

Electrical verification via 20 pF dummy capsule on the APx555: Why standard 5V USB mics clip at 500 mVrms, while our ATD5-T with internal DC/DC converter handles 10 Vrms (28 Vpp).

Verdict: +26 dB higher headroom on the ATD5-T (#6510).
Read Full Lab Report with APx555 Plots →
5. Master for All Calibrations • Traceable Reference Chain

The Core of Our Calibration Chain: APx555 + Spektra SQ-4.2 Pressure Chamber + Class 1 Standards

Every factory calibration, every individually measured Behringer ECM8000, every MiniDSP UMIK-2 we ship, and every rental instrument in our Dresden fleet passes through our automated test bench powered by the Audio Precision APx555, the Spektra SQ-4.2 precision pressure chamber, and reference standards from G.R.A.S. (40AE / 40HL), Brüel & Kjær (4231), and Norsonic.

#S102 • Spektra SQ-4.2

Microphone calibration frequency response

Acoustic frequency response and sensitivity calibration (30 Hz to 16 kHz) for 1/2" and 1/4" measurement microphones (LEMO, ICP/IEPE, P48V) using the APx555 and Spektra SQ-4.2, including PDF report and raw data.

150,00 € Net
178,50 € incl. 19% VAT • net plus shipping
Open Calibration Portal →
#S101 • 20 pF Dummy

microphone calibration self noise

Determination of acoustic and electrical self-noise (via 20 pF dummy capsule on the APx555) in 1/3-octave bands as well as A- and CCIR-weighted.

150,00 € Net
178,50 € incl. 19% VAT • net plus shipping
Open Low-Noise Mics Portal →
#6510 • 10 Vrms / 28 Vpp

NEW: module based USB measurement Microphone 1/2" ATD5-T

Modular 1/2" USB measurement microphone system (15 dB(A) to 143 dB peak, 10 Vrms max input), engineered and individually verified on our APx555.

825,00 € Net
981,75 € incl. 19% VAT • net plus shipping
Open USB Microphones Portal → High-SPL Portal (140–186 dB) →
#1999 • 192 kHz / 64-Bit FFT

software bundle for measuring electric systems

The cost-effective PC measurement suite for your own workbench: 2-channel analysis up to 192 kHz / 24-bit, frequency and level sweeps, THD+N, IMD, and 64-bit FFT.

399,00 € Net
474,81 € incl. 19% VAT • net plus shipping
Open WinAudioMLS Portal →
6. Video Lab • Real APx555 Bench Tests & HPSA Animations

The Audio Precision APx555 in Action

Watch our laboratory bench tests and procedurally computed DSP visualizations explaining the APx555 architecture (GDPR-compliant 2-click loading via youtube-nocookie.com).

Load YouTube Video (2-Click Privacy)
A connection to youtube-nocookie.com is established only after you click.
01:51 min • HPSA Architektur

APx555 Architecture: How the Analyzer Reaches Below -120 dB THD+N

Mathematical visualization of the dynamic range paradox: Analog twin-T notch filter, 1000x residual amplification, and dual-ADC reconstruction.

Load YouTube Video (2-Click Privacy)
A connection to youtube-nocookie.com is established only after you click.
17:46 min • Labor-Messung

APx555 Deep Dive: Ultra-Low THD Measurements (Analog vs. Digital Generator)

Live bench test on the APx555: Comparing the DAC and analog RC generator down to -150 dB in a 64k FFT spectrum and explaining internal gain relay thresholds.

Load YouTube Video (2-Click Privacy)
A connection to youtube-nocookie.com is established only after you click.
18:03 min • Prüfstand-Test

€30 Mic Booster on a €50,000 Analyzer: Klark Teknik CM-1 Tested on the APx555

Reference measurement at 50 µV input level: Verifying 26 dB gain, 6 dB noise floor improvement, and 80 kHz bandwidth.

Load YouTube Video (2-Click Privacy)
A connection to youtube-nocookie.com is established only after you click.
12:25 min • Vergleichstest

€5 Import Mic Booster vs. Klark Teknik CM-1 on the Audio Precision APx555

Metrological proof of 7 kHz parasitic oscillation, 50 Hz mains hum, and -45 dB THD+N on ultra-cheap preamplifiers.

Load YouTube Video (2-Click Privacy)
A connection to youtube-nocookie.com is established only after you click.
03:18 min • Hörtest vs. APx555

THD Listening Test: What Can Human Ears Hear Compared to the APx555?

Systematic injection of harmonic distortion from 10 % down to 0.1 % compared against the 0.0001 % measurement floor of the APx555.

FAQ • AUDIO PRECISION APx555

Frequently Asked Questions About Our APx555 Test Lab & Plugins

Why does the Audio Precision APx555 feature an analog RC oscillator alongside its digital DAC? ▼

Every digital D/A converter is physically limited to roughly -110 dB to -115 dB THD+N due to quantization steps, clock jitter, and reconstruction filtering. By contrast, the microprocessor-controlled analog RC oscillator inside the APx555 generates a continuous-time sine wave with a typical residual THD+N below -120 dB (0.0001 % at 1 kHz and 2.0 Vrms). For complex stimuli such as multitone or logarithmic sweeps, the unit switches to its digital generator.

How does a B2B contract measurement (#9012) or remote session (#9012DIY) work? ▼

With the turnkey service (#9012), you ship your Device Under Test (DUT) to Dresden. We measure up to 5 parameters across 2 channels on the APx555 and return your hardware with a confidential PDF report. With the remote option (#9012DIY), we wire your DUT to the APx555 per your instructions and grant you 1 hour (or more) of live TeamViewer access so you can operate the APx software yourself.

Which Audio Precision analyzers are compatible with the psychoacoustic plugins (#75003 and #75004)? ▼

Our psychoacoustic plugins integrate directly into the Sequence Mode of the Audio Precision APx software suite. They support all analyzers in the APx family: APx555, APx515, APx525, APx585, and APx Flex for ASIO-compatible audio interfaces.

Why do you also test USB measurement microphones like the MiniDSP UMIK-2 or ATD5-T electrically on the APx555? ▼

A purely acoustic pressure-chamber calibration only verifies the signal path at a single moderate reference level (e.g., 94 dB SPL). By replacing the microphone capsule with a precision 20 pF dummy capacitor and feeding the preamplifier directly from the APx555, we determine exact electrical self-noise, THD+N, overload ceiling (500 mVrms on UMIK-2 vs. 10 Vrms on ATD5-T), and frequency response up to 192 kHz sample rate.

Can existing .wav recordings captured on an APx555 be post-processed in Akulap? ▼

Yes. Our Akulap WAV analysis viewer (#3303023) and full Akulap suites read the embedded calibration metadata from Audio Precision APx WAV files (as well as NTi XL2 and Svantek files) directly, enabling post-processing per IEC 61672, DIN 45680, 1/3-octave RTA, and psychoacoustics at sample rates up to 1 MHz.